A low-cost, easy to use, fully integrated system for the calibration of dew-point and relative humidity sensors from 2 to 90% relative humidity, -30 to +20°Cdp.
Description
23 Recent ViewsIn Stock
The HG1 is configured to operate as a stand alone calibrator or to operate from a dry compressed air supply enabling easy on-site calibrations. The calibrator features an integrated chilled mirror reference instrument to enable the operator to perform calibrations that are traceable to national standards. The chamber cover can be customised to suit the transmitters and sensors being calibrated.
Product Features:
Simple operation and maintenance
Range 2 to 90% RH, -30 to +20°C dew point
Built in precision chilled mirror reference instrument
Reference accuracy ±0.2°Cdp
Transportable to allow on-site calibrations
Technical Specification
Model
HG1
Calibration Range
2 to 90% RH (-30 to +20°Cdp) @ 21°C ambient
Reference Accuracy
Typically ±2% of reading (% relative humidity), 0.2°Cdp, 0.1°C ambient temperature (with Optidew reference hygrometer)
Calibration
Through traceable calibration of integrated Optidew reference hygrometer, to NPL and NIST
Operating Temperature
+10 to +35°C ambient
Power
90 to 120 V AC @ 60Hz or 220 to 260 V AC @ 50Hz
Calibration Chamber
Steel with gasket seal, 120 x 120 x 250mm (h x w x d)
Overall Dimensions
Painted aluminium case, 305 x 520 x 400mm (h x w x d)
To contact AMS Instrumentation & Calibration about Michell Instruments Michell Humidity Calibrator | HG1 use Get a quote.
The HG1 is configured to operate as a stand alone calibrator or to operate from a dry compressed air supply enabling easy on-site calibrations. The calibrator features an integrated chilled mirror reference instrument to enable the operator to perform calibrations that are traceable to national standards. The chamber cover can be customised to suit the transmitters and sensors being calibrated.
Product Features:
Simple operation and maintenance
Range 2 to 90% RH, -30 to +20°C dew point
Built in precision chilled mirror reference instrument
Reference accuracy ±0.2°Cdp
Transportable to allow on-site calibrations
Technical Specification
Model
HG1
Calibration Range
2 to 90% RH (-30 to +20°Cdp) @ 21°C ambient
Reference Accuracy
Typically ±2% of reading (% relative humidity), 0.2°Cdp, 0.1°C ambient temperature (with Optidew reference hygrometer)
Calibration
Through traceable calibration of integrated Optidew reference hygrometer, to NPL and NIST
Operating Temperature
+10 to +35°C ambient
Power
90 to 120 V AC @ 60Hz or 220 to 260 V AC @ 50Hz
Calibration Chamber
Steel with gasket seal, 120 x 120 x 250mm (h x w x d)
Overall Dimensions
Painted aluminium case, 305 x 520 x 400mm (h x w x d)
Distribution: Australia, New Zealand, Pacific Islands
Location: Boronia, VIC
Customers: BHP, Santos, Shell, Australian Navy, Coca Cola, CSL, ANSTO, Rio Tinto, FMG, Oil Search
Trade With Confidence. The supplier of this item is approved as a Verified Supplier. Business registration, contact details and locations have been verified. AMS Instrumentation & Calibration are also proudly displaying these credentials:
Established:
1973
Employees:
21-50
Customers:
BHP, Santos, Shell, Australian Navy, Coca Cola, CSL, ANSTO, Rio Tinto, FMG, Oil Search